●Monitoring display: high-definition digital tube display
●Monitoring parameters: tidal volume, respiratory rate, inspiratory ratio,
peak airway pressure
●Breathing mode: IPPV, MAN
●Respiratory rate: 2 times/minute to 99 times/minute
●Respiration ratio: 1:8~5:1, continuously adjustable
●Tidal volume: 0,50~1500mL (using a turbine-type flow sensor)
Minute ventilation volume: not less than 16L/min
●Peak airway pressure: 0~9.9kPa
Safety performance:
●Maximum safe pressure of anesthesia breathing system: 6kPa●High pressure limit: When the airway pressure exceeds the upper limit
set value, the ventilator will automatically limit it to ensure that the
airway pressure cannot continue to rise, so as to ensure the safety of the
patient.
● Insufficient gas source pressure alarm: when the oxygen supply
pressure is lower than 0.2MPa, there will be a high-priority sound alarm,
the duration is greater than 7 seconds, and the sound intensity is greater
than 60dB (A).
● Airway pressure upper limit alarm: 2KPa ~ 6KPa, continuously
adjustable; when the airway pressure exceeds the preset value, there
will be a sound alarm and screen text display immediately, and the
inspiratory phase will be switched to the expiratory phase.
● Airway pressure lower limit alarm: 0KPa~2KPa, continuously
adjustable; when the airway pressure is lower than the preset value, the
delay is 4s~15s (the alarm delay does not exceed 15s), there is a sound
alarm and screen text display.
● Alarm mute: The alarm mute (except for the alarm of insufficient
oxygen supply) is not more than 120 seconds, and the device will
automatically restore the alarm function when the time comes.
● Alarm recovery: After the alarm condition disappears, the alarm
device automatically stops the alarm.